Malt intended for use in beer brewing or elsewhere in the food industry
While α-amylase is active in a gelatinized solution of amylopectin, the viscosity of the solution is constantly dropping due to the degradation of the starch molecules, which can be tracked using a rotational viscometer. The change in the reciprocal of the specific viscosity serves as a measure for the activity of α-amylase.
Suitable for any wort or mash
The time required for a volumetrically defined amount of fluid (the volume between two marks on the tube of the viscometer) to flow through a capillary of a defined length is measured.
Definition:
1 square meter per second is the kinematic viscosity of a homogenous fluid with a dynamic viscosity of 1 Pa × s and a density 1 kg/m3.
The unit for expressing kinematic viscosity (the relationship of viscosity to density = viscoplasticity) is 1 m2/s = 106 mm2/s
The gelatinization temperature can be determined using a rotary viscometer (e.g., Amylograph or Viscograph, Brabender GmbH & Co. KG, Germany [4] or a Rapid-Visco-Analyser, RVA, Perten Instruments, a PerkinElmer Company, USA [8]).
A suspension consisting of a finely ground sample and water is produced, whose precise mixing ratio should correspond to the analysis protocol for the adjunct in question. However, since for many cereals and pseudocereals no official analysis protocol exists, the initial weight for the adjuncts listed in table 2 has been determined empirically [3].
Once the suspension is prepared it is attempered according to a pre-programmed temperature/time program, and the viscosity is determined on a continuous basis by means of a rotor and a rotary torque measurement (fig. 1). When gelatinization begins, an increase in the viscosity is registered, and the corresponding sample temperature is defined as the gelatinization temperature. The standard evaluation criterion (PT) is a viscosity increase of at least 24 cP (≙ mPas) within six seconds.
Malt intended for use in beer brewing or elsewhere in the food industry
Through the action of endo-β-glucanase on a β-glucan solution, the viscosity of the solution continuously decreases, due to the degradation of the substrate molecules, which is measured using a rotary viscometer. The change in the reciprocal of the specific viscosity is a measure for the activity of endo-β-glucanase.
The method describes how to determine the viscosity of the laboratory wort using various kinds of falling ball viscometers.
Suitable for all worts and mashes
The amount of time required for a special ball to fall while sinking through a glass tube filled with a test fluid between two marks is measured. The precision of this method is increased if the time is measured using a photoelectric beam to an accuracy of 0.01 s, rather than manually with a stop watch.
Determination of the viscosity of beverages
wort, beer, beer-based beverages, NAB, juice, beverages in general
Capillary viscometers measure the kinematic viscosity of Newtonian fluids, unlike falling ball and the rotational viscometers, which directly measure the dynamic viscosity. However, with values for the kinematic viscosity and the density of the test liquid, the dynamic viscosity can be calculated.
The time required for a volumetrically defined amount of fluid (the volume between two marks on the tube of the viscometer) to flow through a capillary of a defined length is measured.
1 square meter per second of a homogenous fluid with a dynamic viscosity of 1 Pa × s and a density 1 kg/m³
The unit for expressing kinematic viscosity (the relationship of viscosity to density = viscoplasticity) is 1 m²/s = 106 mm²/s
The kinematic viscosity is expressed as 1 square meter per second of a homogenous fluid with a dynamic viscosity of 1 Pa × s and a density 1 kg/m³.