The method is suitable for the determination of water vapor volatile aroma compounds in beer.
Volatile aroma compounds are driven out of the sample through steam distillation. The ethanolic distillate is saturated with NaCl. Potassium hydrogen sulfite is added to separate carbonyl groups that might interfere with the analysis. The extraction of the aroma compounds is performed by shaking out with dichloromethane and the phases separated by centrifuging.
This method is suitable for the determination of steam-volatile aroma compounds in wort.
Volatile aroma compounds are driven out of the sample through steam distillation. The ethanol distillate is adjusted to be alkaline and saturated with NaCl. The extraction of the aroma compounds is performed by shaking out with dichloromethane and the phases separated by centrifuging. The organic phase is further concentrated in a stream of nitrogen gas. An ammonia solution is added to remove the acids, because the acids would co-elute, thus preventing quantification of the target substances.
The method is suitable for the determination of steam-volatile aging indicators in beer.
Volatile aging indicator substances are driven out of the sample through steam distillation. The ethanol distillate is adjusted to be alkaline and saturated with NaCl. The extraction of the aroma compounds is performed by shaking out with dichloromethane and the phases separated by centrifuging. The organic phase is further concentrated in a stream of nitrogen gas. Ammonia solution is added to remove the acids, as the acids would coelute, thus preventing the quantification of important substances.