R-205.15.170 [2016-03] β-Glucan Content in Wort – Fluorimetric Method

The wort analysis methods are written, for example, for wort that is produced using the Congress mash method. These methods may also be applied in the analysis of wort which has been produced using a different mash method, particularly the isothermal 65 °C method. However, the reference values listed in the analysis method for Congress wort are not transferable.

read more

Application/Purpose

This method describes the fluorimetric determination of high-molecular weight β-glucans in laboratory worts.

Scope of Application

Suitable for all types of (laboratory) worts

Principle

The fluorochrome Calcofluor forms a complex with high molecular weight β-glucans (molecular weight greater than 5 kDa). Complex formation results in an increase in fluorescence; however, this fluorescence is extremely unstable due to photochemical degradation.

Reproducible measurements for fluorescence and determination of β-glucan are possible through measurement in an automatic analysis system based on flow injection (flow-injection analysis). The apparatus is calibrated using purified barley β-glucan standard solutions.